Multi-parameter
controller
Reference at: http://www.consort.be/
Model: R3610
• R3620 • R3624
1.
Overview:
Control:
-
Two independent channels allow to control any two
functions simultaneously, each with its own temperature compensation.
-
On/off or proportional control.
-
Four solid state relays programmable for high/low level
control, wash program or alarm functions.
-
Allows to stop temporarily the control (relays off)
without holding the measurements.
-
Wash program for an automatic periodical cleaning of the
electrodes.
-
A programmable alarm function prevents the overdosing of
chemicals in the process liquid. Each time a pre-set level is exceeded the
corresponding elay is closed and the alarm timer starts to count down. When
this level is still exceeded after count down of the timer all relays will be
opened and an alarm be given. The system should then be checked thoroughly for
e.g. broken electrodes, interrupted cables, empty vessels, etc...
pH:
-
Multi-point (1...5) calibration for more linearity.
-
Selectable resolution from 0.001 pH to 0.1 pH.
-
Automatic calibration with any of eleven pre-programmed
and five user specified pH buffers.Create your own buffer/temperature tables!
mV:
-
Features mV calibration for accurate ORP measurements.
-
Selectable resolution from 0.1 mV to 1 mV.
Ion (R362x only):
-
Direct concentration measurement.
-
Multi-point (2...5) calibration and an additional blank
correction for measuring low concentrations.
Conductivity:
-
Multi-point (1...3) calibration for more linearity.
-
The 4-pole design reduces considerably the problems of
polarisation and fouling. By utilising four electrodes, no current flows
through the measuring circuit. The AC-current is only applied to the outer pair
of rings allow-ing the inner pair of electrodes to measure the voltage without
any polarisation effects. A 4-pole electrode permits to measure with the
highest degree of accuracy and linearity (R3624 only).
-
An electrode with a typical cell constant of 1 cm-1permits
to measure from 0.01 µS/cm to 200 mS/cm in five ranges.
-
An electrode with a typical cell constant of 0.1 cm-1permits
to measure from 0.001 µS/cm to 20 mS/cm in five ranges.
-
An electrode with a typical cell constant of 10 cm-1permits
to measure from 0.1 µS/cm to 2000 mS/cm in five ranges.
-
Automatically selects correct range and frequency.
-
Selectable reference temperature: 20° or 25°C.
-
Automatic calibration with any of three preprogrammed and
three user specified standard solutions.Create your own standard/temp. tables!
-
Accurate low conductivity measurements by eliminating the
capacitive component of the electrode and its cable (avoid the use of long
cables!).
Dissolved oxygen:
-
Operates with a galvanic oxygen electrode requiring no
polarisa-tion time and no zero calibration.
-
Selectable resolution from 0.01 mg/l (0.1%) to 0.1 mg/l
(1%).
-
Rapid air calibration.
-
Manual salinity compensation 0-40.
-
Automatic air pressure compensation 600-1300 hPa.
Temperature:
-
Reads temperatures with 0.1°C resolution.
-
Manual or automatic temperature compensation.
-
Calibrates temperature probe for quality measure-ments.
Inputs:
-
Two inputs for pH, mV, Ion, conductivity or dissolved
oxygen.
-
Two inputs for temperature.
-
R3624:one DIN-8 connector for 4-pole conductivity
electrodes with built-in Pt1000.
Outputs (R362x only):
-
Two programmable 4-20 mA analogue outputs, for
con-nection to a proportional pump, a PLC or an industrial recorder.
-
RS485 interface permits to connect up to 31 contollers in
a network system with a computer.
Display:
-
A large bright LCD screen with white backlight enables to
view all channels simultaneously.
-
The interactive LCD screen provides step by step
instruc-tions in the language of your choice (English, Dutch, French, German).
-
Real-time clock displays time and date.
-
Shows a GLP report on the LCD screen.
Connections:
-
All connections, except for the inputs, can be done
through internal screw terminals.
Cabinet:
-
Robust dust and splash-proof cabinet for wall mounting.
Data-logging (R362x only):
-
Stores up to 12000 values including temperature, time and
date at a programmable interval.
-
An optional powerful data acquisition software for PC +
RS485/RS232 or RS485/Ethernet adaptor is available separately.
-
Can be connected in a network system with up to 31
controllers and a computer.
-
Two-way communication with any computer using RS485.
-
Programmable identification number.
-
Stores minimum/maximum readings for each channel.
Special features:
-
Three year warranty.
-
No electrical interference between electrodes in the same
solution.
-
Galvanic input/output isolation eliminates ground loop
interferences.
-
A first special timer can start a wash program for an
automatic periodical cleaning of the electrodes.
-
A second special timer can give alarm to interrupt the
process control in case of a pump overfeed or an elec-trode failure.
-
Password protection prevents any unauthorised
modifi-cation of the instrument's settings.
GLP:
-
All procedures for a "Good Laboratory Practice"
are on board.
Pre-programmed standards:
-
pH buffers: 1.68, 2.00, 4.00, 4.01, 6.87, 7.00, 9.18,
9.21, 10.01, 12.00, 12.45 (at 25°C).
-
Conductivity: 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm (at
25°C)
2.
Technical Data:
pH:
-
Range: -2...+16 pH
-
Resolution: 0.001 pH
-
Accuracy: 0.1% ± 1 digit
-
Calibration: 1...5 points
-
Buffers: 11 pre-programmed + 5 user specified
-
Temperature compensation: -5...+105°C
-
ISO-pH: 6...8 pH
-
Slope: 80...120%
mV:
-
Range: ±2000 mV
-
Resolution: 0.1 mV
-
Accuracy: 0.1% ± 1 digit
-
Calibration: 1 point
ION (R362x only):
-
Range: 0.01 ng/l...100 g/l
-
Resolution: 3 digits
-
Accuracy: 0.5% ± 1 digit
-
Calibration: 2...5 points + blank
CONDUCTIVITY:
-
Range (cc
dependent): 0...2000 mS/cm
-
Resolution (cc dependent): 0.001 µS/cm
-
Accuracy: 0.5% f.s. of range
-
Calibration: 1...3 points
-
Standards: 3 pre-programmed + 3 user specified
-
Cell constant (cc) 0.1/1/10 cm-1±30%
-
Temperature compensation: -5...+105°C
-
Reference temperature: 20° or 25°C
-
Temperature coefficient natural waters (EN27888)
-
Capacitive compensation
DISSOLVED OXYGEN:
-
Range: 0...60 mg/l (0...600%)
-
Resolution: 0.01 mg/l (0.1%)
-
Accuracy: 1% ± 1
digit
-
Calibration: 1 point
-
Temperature compensation: 0...50°C
-
Salinity compensation: 0...40
-
Air pressure compensation: 600...1300 hPa
TEMPERATURE:
-
Range: -5...+105°C
-
Resolution: 0.1°C
-
Accuracy: 0.1°C
-
Calibration: 1 point
AIR PRESSURE:
-
Range: 600...1300 hPa
-
Calibration: 1 point
CHANNELS:
-
Measurement: 2
-
Temperature: 2
INPUTS:
-
Measurement: 2
BNC, 1012Ω
-
Temperature: 2 BNC, for Pt1000
-
Four-pole conductivity cell 1 DIN-8 (R3624 only)
CALIBRATION:
-
Reminder: 0...999 h
-
GLP
CONTROL:
-
On/Off
-
Proportional
-
Wash program
-
Alarm timer
DISPLAY:
-
LCD: 240x64 pixels
-
White backlight
-
Real time clock
ANALOG OUTPUTS (R362x only):
-
Two outputs:
4...20 mA, max. 300 Ω load
COMMUNICATION:
-
RS485, baud rate
300...19200 b/s
-
Computer via RS485/RS232 interface
DATA-LOGGING (R362x
only):
-
Data sets 12000 +
°C/date/time
-
Modes all
-
Interval: 1 s...4 h
RELAY OUTPUT:
-
Four relays 4
solid state
-
Voltage 12...250 VAC/ min. 1 mA/ max. 1 A
SECURITY:
-
Identification
number
-
Password protection
AMBIENT CONDITIONS:
-
Temperature: 0...40°C
-
Humidity: 0...95%, non condensing
POWER SUPPLY: Mains 210...250 VAC, 50/60 Hz
DIMENSIONS: WxDxH 28x17x6 cm
WEIGHT: Meter 1.3 kg
3.
Order Info:
|
Item
|
Code
|
Description
|
|
01
|
R3610
|
Controller for pH/mV/ion/conductivity/dissolved oxygen
|
|
02
|
R3620
|
Controller for pH/mV/ion/conductivity/dissolved oxygen
|
|
03
|
R3624
|
Controller for pH/mV/ion/conductivity/dissolved oxygen
|
|
03
|
A2002
|
Data acquisition software DIS-2 + RS232/RS485 adaptor
(optional)
|
|
04
|
A2003
|
Data acquisition software DIS-2 + RS485/Ethernet
adaptor (optional)
|